Protecting the provider, the patient and the environment
MIC Dual Chamber (Model A2000S)
- Consists of two independent gas tight chambers
- Can operate one chamber at negative pressure and one chamber at positive pressure
- Can operate without a negative pressure clean room and without any venting
- 112"L x 24"D x 78"H
- Adjustable height
- 700 lb (320 kg)
- Body – 316L stainless steel
- Viewing portion – Chemical- and shatter-resistant plastic
- Oval glove ports are standard
- Lighting – recessed and sealed from work chamber
- Filtration – HEPA filter 99.99% effective with particles
of greater than .5 micron. Air flow enters and exits
chamber through HEPA filters (ISO Class 5)
- Positive and negative pressure available
- Hazardous and non-hazardous solution within one unit
- Electrical Requirements – MIC Dual requires a standard 120V
outlet dual receptacle. The MIC has a factional horsepower centrifugal blower
and light wired into a single cord with standard three-prong plug for each side
of the workstation resulting in two cords. The power requirement is less than 3 amps.
The lighting source is mounted exterior to the interior of the workstation. The
unit does not require a dedicated circuit. There is no heat load or heat transfer
between the unit and the ambient room.
- Rolls through standard pedestrian doorway
- Self-contained, locking casters
- Smooth surfaces facilitate cleaning
- ISO Class 5 conditions maintained by controlled entrance and exit of product and personnel. Product enters through air lock. Personnel work through glove ports without entering the interior.
- Two-piece glove/sleeve design for quick change
- Filtration and Air Handling Systems
- Both entering and discharged air from the chamber shall be HEPA filtered. The quality of filtration provided by the HEPA filters shall be certified to be 99.99% efficient for particles of .3 micron or larger. The internal environment shall meet ISO class 5 conditions.
- Internal filter faces shall be covered with a membrane allowing the surface to be cleaned
- The air handling system does not require venting. If vented the vent is a four inch discharge which is typically attached from the outside vent to the MIC with flex. If vented the vent should be a four (4) inch round duct extending six inches below the ceiling. The duct should deliver 135 cfm at the discharge in the room. Location should be at the center of the discharge filter box which is located on the right side of the MIC (facing the unit) positioning is not critical as the flex connection will compensate for minor location variance. If the duct is servicing a MIC dual chamber the airflow should be 250 cfm.
- The system shall be capable of operating under either positive or negative pressure. The pressure control should be changeable after installation.
- An owner’s manual shall be provided. The manual shall describe operation of the workstation, maintenance and troubleshooting.
- A factory test report to be provided by an independent source indicating that the workstation meets ISO class 5 conditions.
- The manufacturer will provide a one-year limited warranty, which will include correction of defects at no cost. The warranty will include both parts and labor.
- Technical procedures for cleaning, sanitizing and certification shall be provided. In-service and start up services shall be provided including training of multiple shifts.