Protecting the provider, the patient and the environment
MIC Single Chemo (Model A1000N)
- Consists of one independent gas tight chamber
- Can operate at either negative or positive pressure
- Can operate without a negative pressure clean room and without any venting
- 56"L x 24"D x 78"H (single airlock)
- Adjustable height
- Body – 316L stainless steel
- Viewing portion – chemical- and shatter-resistant plastic
- Oval glove ports are standard
- Lighting – recessed and sealed from work chamber
- Filtration – HEPA filter 99.99% effective with particles of greater than .5 micron. Air flow enters and exits chamber through HEPA filters (ISO Class 5)
- Rolls through standard pedestrian doorway
- Self-contained, locking casters
- Smooth surfaces facilitate cleaning
- ISO Class 5 conditions maintained by controlled entrance and exit of product and personnel. Product enters through air lock. Personnel work through glove ports without entering the interior.
- Electrical Requirements:
- MIC Single requires a standard 120V outlet single receptacle. The MIC
has a factional horsepower centrifugal blower and light wired into a single
cord with standard three-prong plug. The power requirement is less than 1.5
amps. The lighting source is mounted exterior to the interior of the workstation.
The unit does not require a dedicated circuit. There is no heat load or heat
transfer between the unit and the ambient room.
- Filtration and Air Handling Systems
- Both entering and discharged air from the chamber shall be HEPA filtered. The quality of filtration provided by the HEPA filters shall be certified to be 99.99% efficient for particles of .3 micron or larger. The internal environment shall meet ISO class 5 conditions.
- Internal filter faces shall be covered with a membrane allowing the surface to be cleaned.
- The air handling system does not require venting. If vented the vent is a four inch discharge which is typically attached from the outside vent to the MIC with flex. If vented the vent should be a four (4) inch round duct extending six inches below the ceiling. The duct should deliver 135 cfm at the discharge in the room. Location should be at the center of the discharge filter box which is located on the right side of the MIC (facing the unit) positioning is not critical as the flex connection will compensate for minor location variance. If the duct is servicing a MIC dual chamber the airflow should be 250 cfm.
- The system shall be capable of operating under either positive or negative pressure. The pressure control should be changeable after installation.
- An owner’s manual shall be provided. The manual shall describe operation of the workstation, maintenance and troubleshooting.
- A factory test report to be provided by an independent source indicating that the workstation meets ISO class 5 conditions.
- The manufacturer will provide a one-year limited warranty, which will include correction of defects at no cost. The warranty will include both parts and labor.
- Technical procedures for cleaning, sanitizing and certification shall be provided. In-service and start up services shall be provided including training of multiple shifts.